Remote control of Xantrex / Sorensen XT 20-3 DC power supplies over GPIB (IEEE-488) on Linux with the M9B GPIB option and a Keysight 82357B adapter. Why the XT ignores its DIP address, and how to talk to it anyway.
Scripts and notes for controlling Xantrex (Sorensen / AMETEK) XT 20-3
programmable DC power supplies over GPIB / IEEE-488 from Linux, using the
internal M9B GPIB option board (XM-6H9B, manual TM-9BGP-6HSN) and a
Keysight / Agilent 82357B USB-GPIB adapter with
linux-gpib.
Source code: github.com/feiglein74/xantrex-xt-20-3
Found by logic-analyzer captures of the TLC7210 GPIB controller and reverse-engineering of the master firmware (M9B V1.14, MC68HC11F1):
| Finding | Detail |
|---|---|
| Primary address | Always 1, regardless of the DIP switches |
| Secondary address | The DIPs set secondary-address ranges (DIP value + 3, +5, +12, +24) meant for multi-channel XT variants; a single-output XT 20-3 answers on any SAD 3–30 |
| Listener wake-up | After power-on only the talker works (serial poll returns 0x10). A REN off→on edge plus a write sweep over the secondary addresses enables the listener |
| Command set | Xantrex proprietary, not SCPI: VSET, ISET, OUT ON/OFF, VOUT?, IOUT?, ASTS?, FAULT?, VMAX?, IMAX?, OVSET? verified |
import gpib
dev = gpib.dev(0, 1, 0x60 + 5, gpib.T1s, 1, 0) # PAD 1, SAD 5
gpib.write(dev, b"VSET 5\n")
gpib.write(dev, b"VOUT?\n")
print(gpib.read(dev, 128)) # b'VOUT 4.998\n'
Run xt_listener_hunt.py once after every power-on first — see the
README for the full
quick start, ASTS bit table, rear-panel J7 pinout and recovery procedure.